SPX series intelligent biochemical incubator manual

<

I. Overview
SPX series intelligent biochemical incubator adopts the world's advanced microcomputer technology and programmable control method. The appearance is beautiful and the craft is special. The appearance is generous and the insulation performance is excellent.
● Microcomputer fully automatic control, touch switch, easy to operate.
● It can set or stop continuously, set time and temperature.
● PID output control.
● With power-down memory and power-down time automatic compensation function.
● Constant temperature control system, fast response and high temperature control accuracy.
● Air duct ventilation, soft wind speed and uniform temperature.
● It adopts reflective tempered coated glass, which has good thermal insulation performance and beautiful appearance.
● With over-Temperature Sensor abnormal protection function to ensure the safety of instruments and samples.
● It is suitable for plant seed germination, cultivation and seedling, insects, small animal breeding, wood and building materials performance tests.

SPX series intelligent biochemical incubator not only has excellent control performance and anti-interference ability, but also can choose to contact working mode or regular working mode to meet different animal and plant growth and various kinds. Biological preservation needs

two. Performance
1. Temperature (°C)
Control range: 0-50 ° C (industry standard 10-40 ° C)
Volatility: ≤ ±1.0 °C Deviation: ≤±1.5 °C Unevenness: ≤±1.5 °C
2. Heating time: 0 ° C rises to 40 ° C ≤ 80 minutes
3. Cooling time: 40 °C down to 10 °C ≤ 120 minutes
4. Working environment: Temperature 0-40 ° C, humidity 85% RH or less, no corrosive gas
5. Compressor delay time protection time: 3 minutes
6. Working mode: continuous operation (compressor intermittent operation)
7. Noise: ≤70db
8. Power requirements: 220V (range 187V-246V), 50HZ

Three. Specifications

model Volume L Dimensions (width × depth × height) Temperature control range °C Temperature control accuracy °C
SPX-150 150 594×594×1250 0-50 ±0.5
SPX-250 250 544×544×1690 0-50 ±0.5
SPX-350 350 594×594×1850 0-50 ±0.5
SPX-450 450 643×648×1850 0-50 ±0.8
SPX-600 600 1250×650×1550 0-50 ±0.8
SPX-850 850 1215×648×1850 0-50 ±0.8
SPX-1000 1000 1250×655×1980 0-50 ±0.8

Fourth, the structure icon

Fifth, the method of use

Button function

Setting / query
This button is the setting button. Press and hold the button for three seconds to enter the setting state. The last digit of the temperature window flashes. The user can press the corresponding plus or minus button to set the temperature. Press and hold the plus or minus button to enter the quick connection. Less. Press the setting/query again, the single digit of the time window flashes, the user can set the working time according to the corresponding plus or minus keys. (If you want to work continuously, set the time to 00. In the working state, this window will display. Set the temperature), the user sets each parameter, and press this button again, the setting is successful. In the setting state, if no button is pressed within 30 seconds, the incubator automatically exits the setting state and the setting parameters are invalid.

illumination
The button is the box illumination button, press the button to illuminate the light, and press the button again to illuminate.

timing
This button is a timing button.
1. If there is a set time, if the time indication window displays END, the incubator is turned off. Press once to enter the running state, and the time display window displays the countdown time. (The remaining time is counted in seconds, the maximum setting time is 999 hours. It is the addition and subtraction keys. Press this key to add 1 and subtract 1 to set the operation. If you press and hold continuously to enter continuous addition and subtraction.
2. Instruction window
The indicator window indicates the parameters of the setting and operation:
Set the temperature/time display window When setting the status, indicate the set temperature or time.
The temperature display in the box indicates the temperature inside the box, in degrees Celsius (°C).

Setting example: The user requires 200 hours of work and a temperature of 20 °C.

step
1. After power-on, press the setting/inquiry button for 3 seconds. At this time, the temperature display indication window digital tube starts to flash. If the user presses the plus 1 minus 1 button, the digital tube will automatically add and subtract. When setting the indicator window number When the set value is reached 20, the temperature is set, press the setting/query button again, the time window single digit flashes, the time setting is 200, the time setting is completed, if the time window is 00, it will work continuously, press Set/query key, the setting is completed.

2. After the end of the operation, the time window displays END, press the time button to re-run the incubator.

six. Protective function
1. Thermal resistance short circuit, open circuit protection: When the thermal resistance is short circuited, the temperature will display “Err”, and the meter will automatically stop heating and cooling.

2. Anti-icing protection: When the set temperature is lower than 10 °C, the incubator automatically stops humidification.

3. Over temperature protection: When the measured temperature is higher than the set temperature by 2 ° C or 47 ° C, the heating is automatically stopped.

Seven. Troubleshooting